FEATURES:
For use during the following: Casual surgery, Paediatric surgery, Plastic surgery, Vascular surgery,
Dermatology, Otorhinolaryngology, Urology, Gynaecology, Veterinary, Dentistry, First Aid
- Possibility of use for minimum invasive surgery
- Possibility of use of bipolar tweezers
- Possibility of use of split neutral electrodes
- Planning of the working conditions
- Memorization of the last used formulations
- Activation of the power with pedal and / or handle
- Digital regulation and indication of the output power
- Adjustment of the bright intensity of the indicators
- Adjustment of the intensity of the sonorous issue
This is a radiosurgical unit which is suitable to light and medium surgery
This unit, by means of selection of the units performance, allows pure cut (CUT), coagulated cut (BLEND),
superficial coagulation (COAG FORCED), deep coagulation, in absence of carbonization, (COAG SOFT)
and, with a special adapter, bipolar coagulation (BIPOLAR).
The digital reading of the delivered power and the operative functions, monitored by microcontroller,
guarantees the absolute reliability of working conditions.
The ability to control the output functions, as well as the output power level by the handle, allows
adjustment of the unit during a surgical operation, without turning the surgeons attention away from the
surgical field.
This unit allows highly professional surgery, thanks to the ergonomic and safety solutions utilized. The unit
is continually monitoring the patients contact with both of the single neutral electrodes as well as the split
neutral electrode. An alarm will sound if contact is lost.
SPECIFICATIONS:
| MODEL |
80 |
120 |
160 |
| Maximum Output power CUT |
80 W - 250 ohms |
120 W - 250 ohms |
160 W - 250 ohms |
| Maximum Output power BLEND |
60 W - 250 ohms |
90 W 200 ohms |
120 W - 200 ohms |
| Maximum Output power COAG FORCED |
50 w - 150 ohms |
80 W - 150 ohms |
100 W - 150 ohms |
| Maximum Output power COAG SOFT |
40 W - 100 ohms |
60 W - 100 ohms |
80 W - 100 ohms |
| Maximum Output power BIPOLAR |
30 W - 100 ohms |
40 W - 100 ohms |
60 W - 100 ohms |
| Working Frequency |
600KHz |
600KHz |
600KHz |
| Patient Circuit |
F |
F |
F |
| Selectable Power |
115 - 230 V /AC |
115 - 230 V / AC |
115 - 230 V /AC |
| Mains Frequency |
50 - 60 Hz |
50 - 60 Hz |
50 - 60 Hz |
| Electrical Input Power |
230VA |
300VA |
350VA |
| Size H x L x D mm |
260 x 110 x 265 |
260 x 110 x 265 |
260 x 110 x 265 |
| Weight Kg |
7 |
7 |
7 |
| Control |
Plate / patient monitoring |
|
|
| |
Delivery time monitoring |
|
|
| |
Output power monitoring |
|
|
| |
Self analysis |
|
|
| Standard Accessories |
Description |
|
Code |
| |
Power supply cable - 2m |
|
- 00100.03 |
| |
Cable for connection to neutral plate |
|
- 00404.06 |
| |
Steel Neutral electrode |
|
- 5365 |
| |
Kit of assorted electrodes (10 pcs) |
|
- 00500.00 |
| |
Reusable handle with switches |
|
- 2835P |
| |
Foot switch |
|
- 00300.00 |
| |
Manual of instructions |
|
- MA003 |
| Safety Standards |
EN60601-1 (1997) |
|
|
| |
EN60601-2 (1995) |
|
|
| |
EN60601-2-2 (1998) |
|
|
| |
EN60601-1-4 (1998) |
|
|
| |
Electrical Class: I CF |
|
|
| |
MDD 93/42/CEE Class: IIb |
|
|
| |
|
|
|
|